Colorimetric CO2 detectors and newborns

Colorimetric CO2 detectors may fail to indicate successful tracheal tube placement in adults in certain circumstances, such as low cardiac output states, and waveform capnography is considered the gold standard. We now have data that demonstrate their inadequacy for neonatal intubation. Ideally, waveform devices should be used by all professionals who intubate patients – from paramedics to neonatologists.

AIM: Clinical assessment and end-tidal CO(2) (ETCO(2)) detectors are routinely used to verify endotracheal tube (ETT) placement. However, ETCO(2) detectors may mislead clinicians by failing to identify correct placement under a variety of conditions. A flow sensor measures gas flow in and out of an ETT. We reviewed video recordings of neonatal resuscitations to compare a colorimetric CO(2) detector (Pedi-Cap®) with flow sensor recordings for assessing ETT placement.
METHODS: We reviewed recordings of infants <32 weeks gestation born between February 2007 and January 2010. Airway pressures and gas flow were recorded with a respiratory function monitor. Video recording were used (i) to identify infants who were intubated in the delivery room and (ii) to observe colour change of the ETCO(2) detector. Flow sensor recordings were used to confirm whether the tube was in the trachea or not. RESULTS: Of the 210 infants recorded, 44 infants were intubated in the delivery room. Data from 77 intubation attempts were analysed. In 35 intubations of 20 infants both a PediCap® and flow sensor were available for analysis. In 21 (60%) intubations, both methods correctly identified successful ETT placement and in 3 (9%) both indicated the ETT was not in the trachea. In the remaining 11 (31%) intubations the PediCap® failed to change colour despite the flow wave indicating correct ETT placement.
CONCLUSION: Colorimetric CO(2) detectors may mislead clinicians intubating very preterm infants in the delivery room. They may fail to change colour in spite of correct tube placement in up to one third of the cases.

Assessment of flow waves and colorimetric CO2 detector for endotracheal tube placement during neonatal resuscitation
Resuscitation. 2011 Mar;82(3):307-12